GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 3 - System setup
3 - 11
4.
•
Press once on the Release pedal to release the brakes.
•
Press once on the Direction Lock pedal to keep direction lock activated. The direction lock keeps the front wheels from swiveling
and blocking the system inside the narrow transportation box.
5.
Carefully move the instrument out of the Transportation Box, down the ramp, with rear end first.
6.
Remove the clear plastic (wrapped around the scanner) from the unit.
7.
Place all the filling material inside the Transportation Box, close it and store it for possible use in the future.
Table 3-7
Unpacking VIVID E9 from the Wooden Box (cont’d) sheet 2 of 2
Step
Task
Transport notice
GA314329-02
NOTICE
Moving the unit in and out of the transportation box
The unit has brakes on all wheels, but direction lock only on the
front wheels. The wheels position when moving the system into
its transportation box are therefore vital. If the wheels are
swiveling when the system is inside the transportation box, it
may jam the system inside the box. The system must be moved
backwards both into and out of the transportation box.
Moving the system into the transportation box
1. Align the front wheel under the front end of the system as
shown in Figure 1.
2. Push the direction lock pedal to lock the front wheels in this
direction. Check that they are locked.
3. Push the Top Console to its lowest and most backward
locked position.
4. Pull the system into the box with the rear end first.
Moving the system out of the transportation box
1. Unlock brakes but keep direction lock activated. The
direction lock keeps the front wheels in position, and secures
the direction stability when the system is rolled out of the
transportation box.
2. Pull the system out of the box with rear end first.
Figure 1: Brakes and direction locks
Figure 2: Console and Wheels position
Front Wheels
Rear Wheels
Direction Lock
Brake
Brakes
Transportation box
CAUTION
Have two people available to unpack the system.
Attempts to move the system considerable
distances or on an incline by one person could
result in injury or damage or both.
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......