GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 307
8-14-2-3
BEP6 - Digital Video Recorder removal procedure
Follow these steps to remove the Digital Video Recorder from a BEP6:
1.) Locate the DVR Board inside the BEP6.
2.) Unscrew the fixing screws from the connector at top of the DVR.
-
The flex between the Graphics Adapter and the DVR board uses finger screws.
-
The flex from the motherboard is fixed with Hex screws.
3.) Disconnect the audio jack and the two flex cables where they attach to the DVR Board.
4.) Disconnect the DVR to SATA Cable.
5.) Remove the fixing screw for the DVR board.
6.) Pull out the DVR board.
Next:
•
8-14-2-5 "BEP6 - Digital Video Recorder installation procedure" on page 8-309.
Figure 8-254 DVR board position in BEP6
DVR
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......