GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 53
8-5-9-3
Rear Cover installation
Follow these steps to install the Rear Cover:
1.) Position the lower edge of the Rear Cover into place on the rear of the VIVID E9 frame.
2.) Tilt the top edge of the Rear Cover toward the VIVID E9 frame.
Be sure the Z mechanism’s manual release handle extends through the air vent on the Rear Cover.
3.) Position the Rear Cover into place.
4.) Gently pull up on the Z mechanism’s manual release handle to confirm proper position through the
Rear Cover air vents.
5.) Install the four screws, two on each side. Install the two lower screws first, then the upper screws.
6.) Inspect the Filter, and clean if necessary.
7.) Install the Filter.
8.) Install the Filter Cover.
9.) Install the Rear Bumper.
10.)Install the Side Covers.
Figure 8-55 Z-lever in upper air ventilation slot
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......