Chapter 8 - Replacement procedures
8 - 111
8-6-9-2 Remove the Main Controller Board (cont’d)
4.) Disconnect three (3x) cables:
-
Disconnect the High Voltage cable (upper, left side in
-
Disconnect the black connector (press on the “button” on the plug and at the same time pull
it out of the connector).
-
Disconnect the connector (right side in
).
5.) Disconnect the plug on the black cable from the connector on the Main Controller board.
NOTE:
The Main Controller Board does not come with the High Voltage Inverter Board, BUT it does
come with the USB Video Board.
6.) If you are going to change the Main Controller board, you may want to remove the High Voltage
Inverter board now. (See:
8-6-11 "High Voltage Inverter Board with Cable replacement" on
.) If not, continue with the next steps.