VIVID E9 S
ERVICE
M
ANUAL
8 - 290
Section 8-11 - Back End Processor (BEP) parts replacement
8-11-15
BEP6 to GFI and Card Rack Backplane Cable replacement
8-11-15-1
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the
Mains Power Cable
from the wall outlet.
3.) Disconnect
all probes
.
4.) Remove the
Left Side Cover
.
5.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.