Chapter 8 - Replacement procedures
8 - 173
8-6-37-2
Button IF Board removal procedure
1.) Disconnect all the five (5x) connectors from the board.
2.) Push the TABs away to release the board.
3.) Remove the board.
8-6-37-3
Button IF Board installation procedure
1.) Install the Button IF Board into the UI Frame, Lower by pressing it under the small hooks.
2.) Ensure the XYZ Buttons Frame cable runs along the Lower Op Panel cable channel so that the
cable is not pinched when the Palm Rest is replaced (see Figure 8-117).
3.) Install the Handle Left Top and Handle Right Top.
4.) Replace the Palm Rest.
5.) Replace the Lower Operator Panel.
6.) Replace the Upper OP Panel/Touch Panel assembly.
Figure 8-179 Button IF Board