GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
5 - 86
Section 5-10 - Power distribution
5-10-4
Power Down Sequence description
5-10-4-1
Overview
There are three possible scenarios for Power Down of the unit:
•
Power Down
•
Forced Power Down
•
Power Loss
Each of the scenarios are described below.
5-10-4-2
Power Down
Press the ON/OFF button (for a short time)
1.) BEP detects the contact of Power (ON/OFF) switch.
2.) PSON_N goes high. This trigs the Main Power Supply to shut down the output voltages.
3.) Controller ACFAIL_N output signal goes low.
4.) Controller TS_OK output signal goes low.
5.) Controller turns OFF the TS (Transmit) voltages.
6.) Controller turns OFF the PMX (Probe MUX) voltages.
7.) Controller turns OFF vo/-15, +/-6, +11V, +24V.
8.) ACFAIL_N output signal goes high.
9.) Controller turns OFF +48V.
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......