GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 192
Section 8-8 - Main Console parts replacement
3.
Filter Locations
A. Rear Filter
- Remove the Cover and
then remove the filter.
B. Bottom Filter
- Remove the Filter
Assembly by lowering the handle.
NOTE: The handle for the bottom
filter is located in the same location for
both Filter Assemblies.
4.
Clean the Rear Filter after it is removed
by removing excess lint or dust from the
soiled side; or vacuum if necessary.
Table 8-5
Removing and cleaning filters, (cont’d) sheet 2 of 3
Steps
Corresponding Graphic