Preparations
When preparing for the replacement, you must perform the following steps:
1.) Record the VIVID E9’s
TCPIP settings
and installed
Option strings
.
2.) Export the
Patient Archive images
to a server or to external media.
3.) Back up the
Report Archive
,
User-Defined Configuration
(
Customer Presets
), and
Service
settings.
4.) If possible, wipe the HDD partitions as described here: Reload the system software from DVD.
Select A: to overwrite all content on HDD.
5.) Power down the VIVID E9.
6.) Disconnect the
Mains Power Cable
from the wall outlet.
7.) Disconnect
all probes
.
8.) Remove the
Side Covers
.
9.) Remove the
Top Cover
.
10.)Remove the
B/W Printer
.
-
Disconnect the
Power Cable
from the
rear of the printer
.
-
Disconnect the
signal cable
from the
I/O
panel.
-
Loosen the printer bracket wing nut.
-
Slide the printer (with signal cable) forward, out of the VIVID E9.
11.)For easier access in the next steps, remove the VIVID E9’s
Front Cover
.