GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 14
Section 8-4 - Reloading the software
8-4-9-3
Initiate software reload from the Recovery Console
Ensure that the VIVID E9 is powered down.
1.) Depress the
On/Off
button on the
Operator Panel
. The VIVID E9 starts.
2.) Press the
Esc
button, on the alphanumeric keyboard, multiple times until the
GRUB
menu is
displayed on the screen.
3.) On the
Grub
menu, select:
Recovery Console
.
The GE Healthcare Recovery Console is displayed.
4.) Call the
Online Center
(
OLC
).
Tell
OLC
that you need a “
Challenge Code
”, and on request, you must read the
Challenge String
on the screen.
OLC
will return a password that you need to type in as the
Response
.
5.) Enter the
Response Code
(
password
) you got from
OLC
.
Figure 8-2 GRUB menu
Figure 8-3 GE Healthcare Recovery Console (Example)
GNU GRUB version 0.97 (638K lower / 2095879K upper memory)
___________________________________________________________
Windows
Recovery Console
Use the | and | keys to select which entry is highlighted.
Press enter to boot the selected OS or 'p' to enter a
password to unlock the next set of features.
USE THE ARROW DOWN KEY
TO SELECT RECOVERY
CONSOLE, THEN PRESS
ENTER.
Enter the password (from OLC).
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......