Section 8-4 - Reloading the software
8-4-10-5 Press A. Install System Software. (Erase all partititions) (cont’d)
To be able to copy anything into the new partititions on the HDD, a reboot is required. You are promped
to press a key to initiate the reboot.
NOTE:
Let the DVD remain in the DVD tray.
3.) Press any key to continue.
The VIVID E9 will reboot.
The monitor may display the message “No sync” for a few seconds before the
BIOS Boot Screen is displayed (see next page).
Figure 8-9 Installing Empty P-scheme onto the HDD
INSTALLING AN EMPTY P-SCHEME
A RESTART IS
REQUIRED
The system needs to be rebooted in order to continue the installation.
Please let the DVD remain in the CD tray, and press any key to continue.