Chapter 8 - Replacement procedures
8 - 285
8-11-11
DVI to Samtec Jumper (Flex) replacement
The DVI to Samtec Jumper (Flex) is included in the BEP6.X Cable Kit - Spare Part,
Part Number 5433408-80.
8-11-11-1
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes.
4.) Remove the Left Side Cover.
5.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.
6.)
Related information:
•
4-2-2 "Power shut down" on page 4-8.
•
8-5-2 "Side Covers replacement" on page 8-38.
8-11-11-2
DVI to Samtec Jumper (Flex) removal
The number of cables in use, depends on the configuration:
•
4D with DVR: one cable is used.
•
4D without DVR: one cable is used.
•
2D with DVR: two cables are used.
•
2D without DVR does not use this cable.
Follow these steps to remove the
DVI to Samtec Jumper (Flex)
:
1.) Disconnect the cable from CN DV12 (on the motherboard).
2.) Repeat the step above for the second cable, if installed.
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......