GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 226
Section 8-10 - Front End Processor (FEP) / Card Cage parts replacement
8-10-3-2
Front Plane Boards removal procedure
Follow these steps to remove the Front Plane Boards:
NOTICE
Read through the removal and installation steps completely before performing. Perform the steps
carefully to avoid damage to the Front Plane Board connectors.
These steps help unseat the Front Plane Boards from the other boards:
1.
Start by pulling out the Relay Board slightly.
2.
Secondly pull the RX Board slightly out.
3.
Repeat once more for the Relay Board.