Chapter 8 - Replacement procedures
8 - 299
8-13-2-2
BEP6 Side I/O Board Assembly removal
Follow these steps to remove the BEP6 Side I/O Board Assembly:
1.) Disconnect the cables connected to the face of the BEP6 Side I/O Board Assembly.
2.) Remove the four screws connecting the BEP6 Side I/O Board Assembly to the BEP.
3.) Pull the BEP6 Side I/O Board Assembly out of the BEP.
Figure 8-249 Position of screws (two are hidden by cables, one is behind the BEP Cover)