GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 193
5.
Clean the Bottom Filter after it is
removed by removing excess lint or dust
from the soiled side. If necessary, use a
clean, soft brush; or vacuum.
6.
Reinstall the rear filter and tuck the
edges of the filter under the Rear
Bumper and Rear Cover. Reinstall Filter
Cover.
7.
Reinstall Bottom Filter by positioning the
filter under the VIVID E9 and placing the
handle into the stowed position. The
filter will be drawen to the filter mounting
magnets.
Table 8-5
Removing and cleaning filters, (cont’d) sheet 3 of 3
Steps