Chapter 7 - Diagnostics / troubleshooting
7 - 15
Section 7-9
Troubleshooting
7-9-1
No Response from VIVID E9 - It’s “Locked”
If the system appears to be locked, please
wait at least 60 seconds
for the watchdog to trap the
situation. This will, if the situation is trapped, bring up a dialog telling that the system is not responding.
1.) Press
Restart
to restart the application. This will save a special debug log. When the system
restarts it will show a dialog where it requests you to save the log.
2.) Enter a descriptive text then press
Save
.
3.) To export the log, press
ALT+D
. This will bring up the same dialog again.
4.) Now, select destination and choose
Export
to write the log-files to selected destination.