
Front Panel Operation
Using Auxiliary Input (AUXIN)
4
4-35
Perform the following procedure to test the overload protection of an oscilloscope:
1.
Connect the calibrator to Channel 1 on the oscilloscope.
2.
Select the voltage type (dc or ac) using the OUT VAL softkey.
3.
Key in the voltage level. (The default value is 5 V.)
4.
If necessary, change the duration. (Refer to the procedure described above.) The
default duration is 10 s.
5.
Check for test results displayed with the UUTTRIP softkey.
4-50.
Using Auxiliary Input (AUXIN)
Using an external generator, the AUX INPUT selection allows you to perform bandwidth
checks up to 2.4 GHz.
See source instrument for
output value.
MODE
auxin
f
f
f
f
f
volt
edge
levsine
marker
wavegen
video
pulse
meas Z
overld
auxin
To access the AUX INPUT menu, press
^
or press the softkey under MODE until
“auxin” appears.
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...