5820A
Operators Manual
6-44
UUT_SENDB
(IEEE-488, RS-232)
Sends binary data as integers to the UUT serial port.
Parameter:
1 to 10 (Integer) Integers to send to the UUT port.
Example:
To send “*IND?,” followed by a new line (ASCII 10), use:
UUTSENDB 42, 73, 68, 78, 63, 10
UUT_SET
(IEEE-488, RS-232)
(UUT Serial Port Set command) Sets the RS-232-C settings for the 5820A Calibrator
rear panel SERIAL 2 TO UUT serial port, which is saved in the 5820A non-volatile
memory. (While saving configuration data in the non-volatile memory, a period of about
2 seconds, the 5820A does not respond to remote commands.) (To set the parameters for
the rear panel SERIAL 1 FROM HOST serial port, see the SP_SET command.) The
factory default values are shown below in bold type. (To return to the factory defaults,
see the FORMAT SETUP command.)
The interface selection sets the command response, with command echo back with
TERM (terminal) and no echo back with COMP (computer).
Parameters:
<baud rate value>, 300, 600, 1200, 2400, 4800, 9600
<flow control>,
XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity>
PNONE (none), PODD (odd),PEVEN (even)
Example: UUT_SET
9600,XON,DBIT8,SBIT1,PNONE
Set the parameters for the rear panel SERIAL 2 TO UUT serial port to the factory default
values.
UUT_SET?
(IEEE-488, RS-232)
(UUT Serial Port Set query) Returns the RS-232-C settings for the 5820A Calibrator rear
panel SERIAL 2 TO UUT serial port. (To return the parameters for the rear panel
SERIAL 1 FROM HOST serial port, see the SP_SET? command.) The factory default
values are shown below in bold type. (To return to the factory defaults, see the FORMAT
SETUP command.)
Responses:
<baud rate value>,
300, 600, 1200, 2400, 4800, 9600
<flow control>,
XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity>
PNONE (none), PODD (odd),PEVEN (even)
Example:
UUT_SET? returns 9600,XON,DBIT8,SBIT1,PNONE
Return the parameters for the rear panel SERIAL 2 TO UUT serial port, as shown, when
set to the factory default values.
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...