Remote Codes
Summary of Commands and Queries
6
6-43
UUT_SEND
(IEEE-488, RS-232)
(Send UUT Data command) Sends data to the UUT serial port in binary block or string
data format over the 5820A Calibrator rear panel SERIAL 2 TO UUT serial port. The
command may be sent over gpib or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation. Include a line feed (RS-232) character to terminate the block data or
End or Identify (EOI) command (IEEE-488).
Parameter:
#2<nn><nn characters string>
(definite length)
#0<character
string> (indefinite
length)
“<character string>“
(character string)
Examples:
UUT_SEND #206F1S2R0
(definite length format)
Send the data F1S2R0 to the UUT in definite length format. The format is #2 (two
numbers follow) 06 (characters follow) F1S2R0 (6 characters).
UUT_SEND #0F1S2R0
(indefinite length format)
Send the data F1S2R0 to the UUT in indefinite length format. The format is #0 then the
characters.
UUT_SEND
"F1S2R0"
(character
string)
Send the data F1S2R0 to the UUT as a character string.
Special Case When the character string sent to a UUT must end in a carriage return (CR)
command or line feed (LF) command or both, you must use the following:
Definite Length Format Follow the instructions above and after the character string add a
command ^J for CR or ^M for LF or both, where ^J means hold down the <Cntl> key
and type the letter J. For example, sending the string REMS in this format with both CR
and LF, you would count 4 characters for REMS and 1 character each for ^J and ^M for a
total of 6 characters. The command would be UUT_SEND #206REMS^J^M then
<enter>. (The ^J and ^M “characters” actually perform the CR and LF functions.)
Indefinite Length Format This format may not be used when a character string requires
CR and LF commands.
Character String Follow the instructions above and after the character string, add a \n for
CR or \r for LF or both, where the alpha character is entered in lower case. For example,
in the terminal mode to send the string REMS in this format with both CR and LF, the
command would be UUT_SEND “REMS\n\r”. In the computer mode where commands
are entered as part of a command string, use double quotes to show embedded quotes. For
example, "uut_send ""REMS\n\r"" ".
The following characters and commands may be implemented as described above:
Carriage
Return
^J
\n
Line
Feed
^M
\r
Tab
Tab
\t
Backspace
^H
\b
Form
Feed
^L
\f
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...