5820A
Operators Manual
4-6
4-8.
Using the Format EEPROM Menu
W
Caution
To avoid losing calibration constants, use with extreme care.
The format nonvolatile memory menu softkeys permanently
erase calibration constants. Pressing ALL or CAL invalidates
the state of calibration of the 5820A.
Pressing FORMAT NV MEM in the utility functions menu opens the following:
Format NV Memory
All CAL SETUP
f
f
f
f
f
All the softkeys in this menu require the rear panel CALIBRATION switch to be in the
ENABLE position. The nonvolatile memory contains calibration constants and dates,
setup parameters, and the user report string. In the case of calibration constants, factory
defaults are the same for all Calibrators. They are not the calibration constants obtained
when the 5820A was calibrated by the factory before shipment. The softkeys are:
•
ALL
Replaces the entire contents of the EEPROM with factory defaults. This would
be used by service personnel after replacing the EEPROM, for example. It is not
required in normal use.
•
CAL
Replaces all calibration constants with factory defaults but leaves all the setup
parameters unchanged. This is also not required in normal use.
•
SETUP
Replaces the setup parameters with factory defaults (Table 4-1) but leaves
the state of calibration unchanged. You do not have to break the calibration sticker
for this operation. Note that remote commands can change the setup parameters. (See
these commands in Chapter 6:
SRQSTR, SPLSTR, *PUD, SP_SET,
UUT_SET
.)
Table 4-1. Factory Defaults for SETUP
Features
Host Connection
gpib (IEEE-488)
Display Brightness*
level 1,0
GPIB Port Address
4
EOL (end of line)
CRLF
Serial Ports
8 bits, 1 stop bit, xon/xoff, parity
none, 9600 baud, wait 30 sec.
Remote I/F
term
Remote commands
(see Chapter 6)
SRQSTR
SRQ: %02x %02x %04x %04x
*PUD string
cleared
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...