4-3.
Warming up the Calibrator
When you turn on the 5820A, allow a warm-up period of at least 30 minutes for the
internal components to stabilize. This ensures that the calibrator meets or exceeds the
specifications listed in Chapter 1.
If you turn the 5820A Calibrator off after warm-up and then on again, allow a warm-up
period of at least twice the length of time it was turned off (maximum of 30 minutes). For
example, if the calibrator is turned off for 10 minutes and then on again, allow a warm-up
period of at least 20 minutes.
4-4.
Using the Softkeys
The five keys just to the right of the
m
(Previous Menu) key are called softkeys.
Softkey key functions appear directly above the key in the Control Display. Pressing a
softkey either changes a value or causes a submenu with new selections to appear on the
Control Display. Softkey menus are arranged in varying levels, as described in Chapter 3.
You can move backwards to previous menu selections by repeatedly pressing
m
.
Although pressing
R
also returns you to the top level menu, it also resets all volatile
settings and return the Calibrator to the voltage mode. Use the
m
key and the mode
keys (
î
, etc.) as your main navigating tool for moving around the menu levels.
4-5.
Using the Setup Menu
Press the front panel
p
key for access to various operations and changeable
parameters. Some parameters are volatile, meaning they will be lost during reset or when
power is turned off. The following descriptions indicates which parameters are
“nonvolatile.”
When you press
p
from the power-up state, the display changes as follows:
CAL
INSTMT
SETUP
UTILITY
FUNCTNS
f
f
f
f
f
This is the primary instrument setup menu. The list below describes submenus available
through each softkey and tells you where you can find further information in the manuals.
•
CAL
(Calibration) Opens the calibration menu. Softkeys in this menu activate
calibration to external standards and calibration check. Another softkey displays
calibration date information and another opens a calibration report menu.
•
INSTMT SETUP
(Instrument Setup) Selects the desired temperature standard and
opens submenus to access Output, Display, and Remote Setups.
•
UTILITY FUNCTNS
(Utility Functions) Allows you to initiate a self-test, format
the nonvolatile memory, and review the instrument configuration software versions
and user report string. Refer to “Utilities Function Menu” for an explanation of these
features.
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...