Remote Operation
Checking 5820A Status
5
5-43
15
14
13
12
11
10
9
0
0
0
SETTLED REMOTE
0
UUTBFUL
7
6
5
4
3
2
HIVOLT MAGCHG TMPCAL
0
0
0
1
0
8
UUTDATA
0
OPER
oq035f.eps
SETTLED
Set to 1 when the output has stabilized to within specification or the TC measurement
has settled and is available.
REMOTE
Set to 1 when the 5820A is under remote control.
UUTBFUL
Set to 1 when data from the UUT port has filled up the UUT buffer.
UUTDATA
Set to 1 when there is data available from the UUT port.
HIVOLT
Set to 1 when the 5820A is programmed to a voltage above 33 Volts.
MAGCHG
Set to 1 when the output magnitude has changed as a result of another change.
This bit is always 0 in the ISR. It changes to 1 only in the ISCR0 and ISCR1 registers.
TMPCAL
Set to 1 when the 5820A is using temporary (non-stored) calibration data.
OPER
Set to 1 when the 5820A is in operate, 0 when it is in standby.
Figure 5-10. Bit Assignments for the ISR, ISCEs and ISCRs
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...