5820A
Operators Manual
1-10
1-11.
General Specifications
Warmup Time
Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time
5 seconds or faster for all functions and ranges
Standard Interfaces
IEEE-488 (GPIB), RS-232
Operating: 0 °C to 50 °C
Calibration (tcal): 15 °C to 35 °C
Temperature
Performance
Storage: -20 °C to 70 °C
Electromagnetic
Compatibility
Designed to operate in Standard Laboratory environments where the Electromagnetic
environment is highly controlled. If used in areas with Electromagnetic fields >1 V/m,
there could be errors in output values. From 80 – 252 MHz, the current output is
susceptible to a field strength of > 0.165 V/M.
Temperature Coefficient
Temperature Coefficient for temperatures outside tcal
±
5 °C is 0.1X/°C of 1-year
specification.
Operating: < 80 % to 30 °C, < 70 % to 40 °C,< 40 % to 50 °C
Relative Humidity
[1]
Storage: < 95 %, noncondensing
Operating: 3,050 m (10,000 ft) maximum
Altitude
Nonoperating: 12,200 m (40,000 ft) maximum
Safety
Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-
C22.2 No. 1010.1-92
Output Impedance
The 5820A is designed to drive both 50
e
and 1 M
e
loads
Analog Low Isolation
20 V
EMC
Complies with EN 61326-1
Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 to 63 Hz
Line Power
Line Voltage Variation:
±
10 % about line voltage setting
Power Consumption
250 VA
Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm (0.6 inch) for feet on
bottom of unit.
Width: 43.2 cm (17 inches), standard rack width
Dimensions
Depth: 47.3 cm (18.6 inches) overall
Weight
20 kg (44 pounds)
Absolute Uncertainty
Definition
The 5820A specifications include stability, temperature coefficient, linearity, line and
load regulation, and the traceability of the external standards used for calibration. You
do not need to add anything to determine the total specification of the 5820A for the
temperature range indicated.
Specification Confidence
Interval
99 %
[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be required.
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...