Remote Codes
Summary of Commands and Queries
6
6-31
REFOUT?
(IEEE-488, RS-232)
(Reference Output query) Returns the present value of the reference which is the output
values of the 5820A the last time a new reference was established with an OUT, NEWREF,
or MULT. This applies only to VOLTAGE and MARKER mode.
Response: (Float)
Reference
amplitude
Example:
REFOUT? returns 0E+00
Return 0 when the output is not being edited.
Example:
REFOUT? returns 2.500000E-01
Return .250 when the output is being edited and the reference is, for example, 250 mV.
REMOTE
(RS-232)
(Remote command) Places the 5820A Calibrator into the remote state. This command
duplicates the IEEE-488 REN (Remote Enable) message. When in the remote state, the
Control Display shows the softkey “REMOTE CONTROL Go to Local.” Pressing this
softkey returns the 5820A Calibrator to local operation If the front panel is locked out, the
Control Display shows the softkey “REMOTE CONTROL LOCAL LOCK OUT.” (See the
LOCKOUT command.) To unlock the front panel, use the LOCAL command, or cycle the
5820A Calibrator power switch.
Parameter: (None)
Example: REMOTE
Place the 5820A Calibrator in the remote state and display this state on the front panel
Control Display with a softkey REMOTE CONTROL.
*RST
(IEEE-488, RS-232, Overlapped)
(Reset Instrument command) Reset the 5820A Calibrator to the power-up state. *RST holds
off execution of subsequent commands until the reset operation is complete. This command is
the same as pressing the front panel R key.
A reset action evokes the following commands and values:
Command
Value
Command
Value
STBY
(No output)
TRIG
OFF
OUT
0 V,0 HZ
OVLD
10 s
DC_OFFSET
0 V
OUT_CHAN
1
VIDEOFMT
NTSC
TRIG_CHAN
5
VIDEOMKR
line 10
OUTPUT
20 mV
TRIGLMT
10 s
SCOPEMODE
V
RANGELCK
OFF
TDPULSE
OFF
MARKER
spike
SCOPE
OFF
WAVEGEN
square
OUT_IMP
Z1M
Leveled Sine goes to “jump” at a rate of 1 MHz. Refer to Chapter 4 for more information.
Changes made to the setup menus that are not saved in memory are discarded on reset.
Response: (None)
Example: *RST
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...