Remote Codes
Summary of Commands and Queries
6
6-39
TLIMIT_D
(IEEE-488, RS-232)
Sets the default OPERATE time limit for the OVERLD mode signal. At the end of the
specified interval, the instrument will automatically to STANDBY.
Parameters: 1 to 60 (seconds)
Example: TLIMIT_D
15
TLIMIT_D?
(IEEE-488, RS-232)
Returns the default overload time limit.
Response: <Integer> Default time limit in seconds.
TMWAVE
(IEEE-488, RS-232, Coupled, Overlapped)
Selects the waveform for MARKER mode.
Parameters: SINE
Sine
wave (2 ns to 15 ns)
SPIKE
Triangular/sawtooth pulse (15 ns to 5s)
SQUARE
Square wave (50 % duty cycle) (4 ns to 5s)
SQ20PCT
Square wave (20 % duty cycle) (85 ns to 5s)
Example: TMWAVE
SPIKE
TMWAVE?
(IEEE-488, RS-232)
Returns the MARKER mode waveform setting.
Parameters: None
Response: <character> (Returns SINE, SPIKE, SQUARE, or SQ20PCT.)
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...