5820A
Operators Manual
5-16
4.
Verify the 5820A Calibrator is powered and in the reset condition (if in doubt, press
o
), then click the Command1 button (below).
oq025f.bmp
5.
Observe the 5820A Calibrator Control Display changes to REMOTE CONTROL
(below).
REMOTE CONTROL
Go to Local
f
f
f
f
f
6.
Click the Command2 button. Observe the 5820A Calibrator Control Display changes
back to the reset condition (below).
(The Command3 button is used for RS-232 UUT port testing later in this chapter.)
CHANGE
to DC
SCOPE Z
1 M
Ω
V/DIV
MENU
TRIG
off
MODE
volt
f
f
f
f
f
7.
Close the program by clicking the top-left corner and Close.
5-10.
Setting up the RS-232 UUT Port for Remote Control
The SERIAL 2 TO UUT serial data port connects a UUT to a PC or terminal via the
5820A Calibrator (Figures 5-1 and 5-2). This “pass-through” configuration eliminates the
requirement for two COM ports at the PC or Terminal. The UUT_* commands (see
Chapter 6) handle the UUT port data flow.
The RS-232 cable length for each port should not exceed 15 meters, although longer
cable lengths are permitted if the load capacitance measured at a connection point
(including signal terminator) does not exceed 2500 pF.
5-11.
RS-232 UUT Port Setup Procedure
Complete the following procedure to set up the SERIAL 2 TO UUT port (defaults in
bold
). The RS-232 parameters you are selecting must match the RS-232 parameters of
the UUT. This port operates independently, whether the 5820A Calibrator is controlled
from the IEEE-488 port or from the RS-232 Host serial port. (See Figures 5-1 and 5-2.)
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...