Front Panel Operation
Calibrating the Pulse and Frequency Response on an Oscilloscope
4
4-19
CHAN 2
CHAN 4
EXT TRIG
EXT TRIG
SOURCE/MEASURE
CHAN 1
CHAN 2
CHAN 3
CHAN 4
CHAN 5
20V PK
MAX
CHAN 1-5
AUX
INPUT
20V PK
MAX
20V PK
MAX
CAT
130V
PK
MAX
SOURCE
30V DC
MAX
MEASURE
5820A
OSCILLOSCOPE
CALIBRATOR • 2GHz
yh041f.eps
Figure 4-2. Tunnel Diode Pulser Connections
4-32.
The Leveled Sine Wave Function
The Leveled Sine Wave (LEVEL SINE) function uses a leveled sine wave, whose
amplitude remains relatively constant over a range of frequencies, to check the
oscilloscope’s bandwidth. When you check your oscilloscope, you change the wave’s
frequency until the amplitude displayed on the oscilloscope drops 30 %, which is the
amplitude that corresponds to the -3 dB point. Default values are 30 mV p-p, 50 kHz.
Содержание 5820A
Страница 10: ...5820A Operators Manual vi ...
Страница 14: ...5820A Operators Manual x ...
Страница 16: ...5820A Operators Manual 1 2 ...
Страница 32: ...5820A Operators Manual 2 2 ...
Страница 40: ......
Страница 144: ...5820A Operators Manual 6 2 ...
Страница 190: ......
Страница 212: ...5820A Operators Manual 7 24 ...
Страница 214: ...5820A Operators Manual 8 2 ...
Страница 230: ...5820A Operators Manual 9 2 ...
Страница 234: ...5820A Operators Manual ...
Страница 244: ...5820A Operators Manual A 10 ...
Страница 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Страница 246: ...5820A Operators Manual B 2 ...
Страница 248: ...5820A Operators Manual B 4 ...
Страница 250: ...5820A Operators Manual C 2 ...
Страница 258: ...5820A Operators Manual E 4 ...
Страница 264: ...5820A Operators Manual F 6 ...
Страница 272: ...5820A Operators Manual 8 ...