100/180 MM DISTRIBUTED GRAPHIC RECORDER: USER GUIDE
User Guide
Page 340
HA029324
Issue 11 Oct 14
B2.2.1 BATTERY REPLACEMENT PROCEDURE
CAUTION
The user must be at the same electrical potential as the instrument housing when accessing the battery.
Note: All battery backed RAM data is lost during battery change unless power to the unit is main-
tained during the change (see
Annex A
for details of stored data).
The battery is of poly-carbonmonofl uoride/lithium construction and should be disposed of according to
local regulations covering this type of battery.
1. Remove the screw securing the access cover to the rear of the instrument, and carefully remove the
cover.
2. The battery can now be replaced,
Access
cover
BR2330
Cover retaining
screw location
Battery
Figure B2.2.1a Access cover location - large frame unit
Figure B2.2.1b Battery location - small frame unit
(Access cover removed)
Notes:
1. Battery type is BR2330 Poly-carbonmonofl uoride/lithium. Available from the recorder manu-
facturer under part number PA261095.
2. Dispose of exhausted batteries in accordance with local regulations regarding this battery
type.