Liquiline System CA80AM
Commissioning
Hauser
65
9.4.2
Basic setup analyzer
For quick commissioning, you can ignore the additional settings for outputs, relays etc. You can
make these settings later in the specific menus (see table below).
‣
Return to measuring mode by pressing and holding the soft key "ESC" for at least one second.
Your analyzer is now operating using your basic settings.
If you want to configure your most important input and output parameters already in the "Basic
setup analyzer", proceed as follows:
‣
Configure the current outputs, relays, limit switches and device diagnostics using the
following submenus.
Path: Menu/Setup/Basic setup analyzer
Function
Options
Info
Device tag
Customized text,
32 characters
Factory setting
Analyzer_serial number
Select any name for your analyzer. Use the TAG name
for example.
Set date
Depends on the format
Editing mode:
DD (day): 01 to 31
MM (month): 01 to 12
YYYY (year): 1970 to 2106
Set time
Depends on the format
Editing mode:
hh (hour): 00 to 23 / 0 am to 12 pm
mm (minutes): 00 to 59
ss (seconds): 00 to 59
Extended setup
See the "Extended setup" section
Bottle insertion
Procedure for inserting the bottles
Bottle selection
Options
• None
• Cleaner
• Standard 1
• Reagent 1
• Reagent 2
It is possible to select multiple bottles.
Make sure that the hoses are connected to the bottles
before you select "Bottle insertion".
Confirm bottles inserted
Connect the hoses to the bottles. The hoses of the
selected bottles are filled.
Measurement SP%0V
See the "Measurement" section
Calibration
See the "Calibration" section
Cleaning
See the "Cleaning" section