Commissioning
Liquiline System CA80AM
68
Hauser
9.5.3
Device status
Icons on the display alert you to special device conditions.
If two or more diagnostics messages occur simultaneously, only the icon for the message
with the highest priority is shown on the display (for the order of priority according to
NAMUR, see the "Changing diagnostic information" section).
9.5.4
Assignment views
"Assignment views" appear as the last function in many sections of the menu, e.g. Channel
assignment view.
You can use this function to see which actuators or functions are connected to an analyzer
channel or a sensor channel.
The assignments appear in hierarchical order.
Icon
Location
Description
Title bar
Diagnostics message "Failure"
Title bar
Diagnostics message "Maintenance request"
Title bar
Diagnostics message "Check"
Title bar
Diagnostics message "Out of specification"
Title bar
Fieldbus or TCP/IP communication active
Title bar
Hold active (for sensors)
At measured value
Hold for the actuator (current output, limit switch etc.) is active
At measured value
1)
1)
Only pH or ORP measurement
An offset has been added to the measured value
At measured value
Measured value in "Bad" or "Alarm" condition
At measured value
Automatic temperature compensation active (for sensors)
At measured value
Manual temperature compensation active (for sensors)
Title bar
Simulation mode active or Memocheck SIM connected
At measured value
The measured value is influenced by a simulated value
At measured value
The displayed measured value is simulated (for sensors)