Diagnostics and troubleshooting
Liquiline System CA80AM
144
Hauser
304 Module data
F
On
On
At least 1 module has incomplete configuration
data
• Check module information
305 Power consumption
F
On
On
Total power consumption is too high
1. Check installation
2. Remove sensors/modules
3. Restart the device
306 Software error
F
On
On
Internal firmware error
Device configuration does not match firmware
• Contact the Service Department
312 Temperature sensor
F
On
On
Temperature sensor PT3 for ambient
temperature measurement is faulty.
• Winter operation regulation not possible
• Sampling and distribution arm blocked to
protect against freezing
1. Deactivate winter operation mode at
Setup/Inputs/Temperature S:3/Winter
operation
2. Replace sensor
360 Cooling/Heating
S
On
Off
Temperature range in the housing exceeded
1. Check installation conditions and the
ambient temperature
2. Replace the FMAB1 module
361 Cooling/Heating
F
On
On
Cooling/heating module defective
Defined temperature range not reached. This
could affect the functionality of the reagents.
1. Check whether the insulation cap is seated
correctly over the reagents
2. Replace cooling/heating module
362 Photometer temp.
F
On
Off
Temperature of the photometer too high
1. Replace photometer
2. Contact the Service Department
363 Photometer temp.
F
On
Off
Temperature of the photometer too low
1. Replace photometer
2. Contact the Service Department
364 Timeout liquidmanag.
F
On
On
The Liquidmanager/dosing dispenser has timed
out.
Possible reasons:
• Photoelectric barrier defective
• Blocking
-> Contact the Service Department
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current