Liquiline System CA80AM
Diagnostics and troubleshooting
Hauser
143
11.6 Overview of diagnostic information
11.6.1 Device-specific, general diagnostic messages
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current
202 Selftest active
F
On
Off
Self-test active
215 Simulation active
C
On
Off
• Simulation enabled
• Simulation can be stopped in the diagnostic
menu or by restarting the device
242 Firmware incomp.
F
On
On
Firmware incompatible
1. Check firmware version
2. Update firmware
243 Firmware failure
F
On
On
Internal device error
1. Update the software
2. Replace displayed assembly
3. Contact the service department and quote
the number displayed
261 Electronics module
F
On
On
Electronics module defective
1. Restart the device
2. Check module
3. Replace module
262 Module connection
F
On
On
Electronics module not communicating
1. Check module
2. Replace electronics module
3. Check the internal connecting cable to the
module
263 Incomp. detected
F
On
On
Incompatibility detected
The device configuration does not match the
configuration.
1. Check device settings
2. Check electronics module type
284 Firmware update
M
On
Off
Firmware update successfully completed
285 Update error
F
On
On
Firmware update failed
Possible reasons:
• Read error
• SD card is faulty
• Incorrect firmware on SD card
1. Repeat update
2. Use another data storage medium with
firmware
3. Check data storage medium for correct
firmware
302 Battery low
M
On
Off
Buffer battery of real time clock is low
It is located on the backplane. The date and time
are lost if the power is interrupted.
• Replace battery