Liquiline System CA80AM
Operation
Hauser
101
Signal type = "PFM"
Max. frequency
1.00 to 1000.00 Hz
Factory setting
1000.00 Hz
Maximum frequency of the PFM output signal
Is to equal the maximum possible upper limit of the
measuring range.
Meas. value format
Options
• #
• #.#
• #.##
• #.###
Factory setting
#.##
Specify the number of decimal places.
Source of data
Options
• None
• Sensor inputs
• Binary inputs
• Controller
• Fieldbus signals
• Mathematical functions
Factory setting
None
Source, whose value should be read out as a frequency
over the binary output.
Measuring value
Source of data = an input
Selection dependent on
"Source of data"
"Inputs" = everything except controllers
Actuator type
Source of data = a controller
Options
• None
• Bipolar
• U
• Unipolar-
Factory setting
None
Determines which controller component the connected
actuator, e.g. the dosing pump, should trigger.
U
Part of the manipulated variable that the controller uses
to increase the process value
Unipolar-
For connected actuators that decrease the controlled
variable
Bipolar
"Split range"
Hold behavior
Options
• Freeze
• Fixed value
• None
Factory setting
Freeze
Freeze
The device freezes the last value.
Fixed value
You define a fixed current value that is transmitted at
the output.
None
A hold does not affect this output.
Hold value
Hold behavior = "Fixed value"
0 to 100 %
Factory setting
0 %
Path: Menu/Setup/Outputs/Binary output x:y
1)
Function
Options
Info