11.12 Systemtest
Temperature
-50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
Set the desired simulation value.
Diagnostics/Systemtest
Function
Options
Info
Analyzer
Sample collector
Is only displayed if a collecting vessel is provided.
Empty sample collector
The sample collecting vessel can be automatically
emptied in this menu.
Start
Stop
Sample preparation 1
(CAT820/CAT860)
Depends on the connected sample preparation system
Cabinet heating
Tests the housing heating.
Cabinet temperature
Read only
Displays the current housing temperature
Mode
Read only
On for 10 minutes
The heating is switched on for 10 minutes.
Off
The heating is switched off.
Automatic
The heating is automatically switched on and off in line
with the housing temperature.
Hose heating filter
Tests the hose heating (filter to pump)
Ambient temperature Read only
Displays the current outside temperature
Mode
Read only
On for 10 minutes
The heating is switched on for 10 minutes.
Off
The heating is switched off.
Automatic
The heating is automatically switched on and off in line
with the outside temperature.
Hose heating analyzer
Tests the hose heating (pump to analyzer).
Ambient temperature Read only
Displays the current outside temperature
Mode
Read only
Diagnostics/Simulation
Function
Options
Info