Maintenance
Liquiline System CA80AM
170
Hauser
1. Rinse the system (see the "Rinsing the system" section).
2. Remove the bottles and the bottle tray.
3. Remove the cover of the carrier board.
4. Select "Menu/Operation/Dispenser replacement/Dispenser selection".
5. Select the dispenser n that you wish to replace, e.g. dispenser 2 and 7 (10 ml) or dispenser
3 and 4 (2.5 ml).
6. Select "Empty dispenser".
7. Open the dispenser holder by pressing both snap-fit hooks together and removing it.
8. Turn the dispenser clockwise to remove it from the liquid manager.
9. Remove the adapter and the dispenser from the dispenser drive. To do so, hold the dispenser
by the black block at the lower end and pull it from the metal pin.
10. Screw a new dispenser onto the liquid manager.
Make sure that the end of the dispenser is parallel to the liquid manager as the dispenser
holder unit cannot be closed otherwise.
Make sure that the dispenser is connected correctly.
11. Slide the dispenser holder over the snap-fit hooks. Ensure that the holder snaps into place.
Make sure that the drain hose D6 is routed behind dispenser 7.
12. Secure the cover and place the bottle tray back into the housing.
13. Connect the bottles to the appropriate covers and hoses.
14. Select "Menu/Operation/Bottle replacement/Bottle insertion/Bottle selection".
15. Select all the bottles and press "OK" to confirm.
16. Then select "Confirm bottles inserted".
17. Select "Menu/Operation/Dispenser replacement/Dispenser selection".
18. Select the dispensers that you replaced and click "OK" to confirm.
19. Select "Reset operating hours counter".
20. The system must be calibrated after a replacement routine. Select "Menu/Setup/Manual
operation/Determine calibration factor".
21. Press the soft key "MODE" and select "Continue automatic mode" or "MODE/Start automatic
mode".