10.6.4 EtherNet/IP
5)
Determine which process values should be output via EtherNet/IP communication.
You can define a maximum of 16 analog device variables (AI):
‣
Define the data source. You can choose from the analyzer, as well as sensor inputs, binary
inputs and controllers.
‣
From the source, select the measured value that should be output.
‣
Define how the device should behave in the event of a "Hold" (for sensors).
‣
For controllers, you also specify the manipulated variable type.
‣
Please note that if you select Hold behavior="Freeze", the system not only flags the status but
also actually "freezes" the measured value.
In addition, you can define 8 digital device variables (DI):
‣
You specify only the data source here. You can choose from relays, binary inputs and limit
switches.
Additional information can be found on the product pages on the internet (SD01293C).
5)
Available from software version 01.05.03