Device description
Liquiline System CA80AM
14
Hauser
3.3
Device architecture
3.3.1
Slot and port assignment
The electronic configuration follows a modular concept:
• Several slots are available for electronics modules. These are referred to as "slots".
• These slots are numbered consecutively in the housing. Slots 0 and 1 are always reserved for
the basic module.
Slot 2 is the analyzer interface module if no other modules are present.
• Each electronics module has one or more inputs and outputs or relays. Here they are all
collectively known as "ports".
• Ports are consecutively numbered per electronics module and are recognized automatically by
the software.
a0021498-de
Fig. 8: Slot and port assignment of hardware and on-screen display
E
85
85
86
86
41
43
42
87
88
97
98
31
31
32
32
SD
Display
1
2
Sensor
0/4 ... 20 mA
Sensor supply
+
+
–
+
+
–
1
2
Service
PK
GY
PK
GY
GY
1
2
–
BN
WH
GN
YE
BN
WH
GN
YE
87
88
97
98
+
–
A
B
+
–
A
B
–
Alarm
HAR
T
Power (internal)
24VDC
24V
12,5V
3,3V