Liquiline System CA80AM
Diagnostics and troubleshooting
Hauser
145
365 Photometer comm.
F
On
On
Photometer not communicating
Possible reasons:
• Incorrect photometer connection
1. Check photometer connection
367 Module connect.
F
On
On
No communication to sample preparation
-> Check connecting cable to sample preparation
system.
370 Internal voltage
F
On
On
Internal 24-V voltage outside the valid range
--> Check supply voltage
373 Electronictemp. high
M
On
Off
Temperature of backplane electronics is high
--> Check ambient temperature and energy
consumption
374 Sensor check
F
On
Off
Sensor measuring sequence counter
No measurement signal from sensor
1. Check sensor, replace if necessary
2. Check sensor connection
380 Firmware failure
F
On
On
Internal software error
1. Update the software
2. Replace backplane
3. Contact the service department and quote
the number displayed
401 Factory reset
F
On
On
Factory reset is performed
406 Param. active
C
Off
Off
Configuration enabled
--> Wait for configuration to be finished
407 Diag. active
C
Off
Off
Diagnostic menu enabled
--> Wait for maintenance to be finished
408 Calibration aborted
M
Off
Off
Calibration was aborted
412 Writing backup
F
On
Off
--> Wait for the write process to be finished
413 Reading backup
F
On
Off
--> Wait
460 Output below limit
S
On
Off
Current output not reached
Measured value is outside the specified current
range
Possible reasons:
• Sensor in air
• Air pockets in assembly
• Incorrect flow to sensor
• Sensor fouled
Measures
1. Check the application
2. Check configuration of current output
3. Clean sensor
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current
Содержание Liquiline System CA80AM