Operation
Liquiline System CA80AM
72
Hauser
10.1.3 Hold settings
Path: Menu/Setup/General settings/Hold settings
Function
Options
Info
Settings automatic Hold
Hold release time
0 to 600 s
Factory setting
0 s
The hold status is maintained for the duration of the
delay time when you switch to measuring mode.
Setup menu
Options
• Disabled
• Enabled
Factory setting
Disabled
Determine if, when the menu is called up, a hold status
should be issued at the current output.
Diagnostics menu
Options
• Disabled
• Enabled
Factory setting
Disabled
Determine if, when the menu is called up, a hold status
should be issued at the current output.
Calibration active
Options
• Disabled
• Enabled
Factory setting
Enabled
Determine if, when the menu is called up, a hold status
should be issued at the current output.
External hold
Options
• None
• EtherNet/IP DO 1 to 8
Factory setting
None