Liquiline System CA80AM
Diagnostics and troubleshooting
Hauser
149
931 Sampling time
M
On
Off
Standard pumping time exceeded
Possible reasons:
• Filter cartridge clogged
• Sample line partially blocked
1. Clean the filter cartridge
2. Clean the sample line
3. Replace the filter or sample line
932 Cleaning failure
F
On
On
Cleaning failed
Possible reasons:
• Suction line blocked or leaking
• Compressed air supply defective
• No inflow of cleaner or sample
1. Check the suction line
2. Check the compressed air supply and hoses
3. Check the cleaner and cleaner pump
4. Check inflow of sample
936 Temperature range
S
On
Off
Outside temperature for sample preparation
system out of specification
1. Check application
2. Check temperature sensor
3. Check configuration of the temperature
sensor, sample preparation system and the
heating
940 Process value
S
On
Off
Measured value out of specification
Uncertain measured value.
1. Change the measuring range
2. Calibrate the system
941 Process value
F
On
On
Measured value out of specification
Invalid measured value.
1. Change the measuring range
2. Calibrate the system
951
-
958
Hold active CH1 ... 8
C
On
Off
Output values and status of the channels are on
hold. Wait until the hold is released again.
969 Modbus Watchdog
S
Off
Off
The device did not receive a Modbus telegram
from the master within the specified time. The
status of Modbus process values received is set to
invalid.
970 Curr. input overload
S
On
On
Current input overloaded
The current input is switched off from 23 mA due
to overload and reactivated automatically when a
normal load is present.
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current