Diagnostics and troubleshooting
Liquiline System CA80AM
146
Hauser
461 Output above limit
S
On
Off
Current output exceeded
Measured value is outside the specified current
range
Possible reasons:
• Sensor in air
• Air pockets in assembly
• Incorrect flow to sensor
• Sensor fouled
Measures
1. Check the application
2. Check configuration of current output
3. Clean sensor
462 Output Deviation
S
On
Off
When the current output is read back, the value
deviates from the target value displayed.
Possible reasons: Current load outside
specification, short-circuit or open current loop,
module defective
1. Check installation of current loop
2. Check module
3. Contact the Service Department
502 No text catalog
F
On
On
--> Contact the Service Department
503 Language change
M
On
Off
Language change failed
--> Contact the Service Department
530 Logbook at 80%
M
On
Off
1. Save the logbook to the SD card and then
delete the logbook in the device
2. Set memory to ring memory
3. Deactivate logbook
531 Logbook full
M
On
Off
532 License error
M
On
Off
--> Contact the Service Department
540 Parameter save fail
M
On
Off
Configuration saving has failed, --> repeat
541 Parameter load ok
M
On
Off
Configuration successfully loaded
542 Parameter load fail
M
On
Off
Configuration loading has failed, --> repeat
543 Parameter load abort
M
On
Off
Configuration loading aborted
544 Parameter reset ok
M
On
Off
Factory default successful
565 Configuration
M
On
Off
Invalid configuration of sample preparation
system
-> In Setup/Sample preparation check the
number of channels used, their mode of operation
and installation method.
-> Check the permitted combinations of sample
preparation and analyzers, see the Operating
Instructions for the sample preparation system.
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current