Maintenance
Liquiline System CA80AM
166
Hauser
12.2 Replacing reagents, standard and cleaning agent
1. Press the soft key "MODE" and select "Manual mode".
2. Wait until all operations have stopped.
3. Remove the bottles that you wish to replace by removing the hoses at the liquid manager.
Use a paper towel to catch any chemicals that may escape and replace them with fresh
reagent, standard solution or cleaning agent.
4. Select "Menu/Operation/Maintenance/Bottle change mode/Bottle removal/Bottle
selection".
5. Select the bottles that you want to remove and click OK to confirm.
6. Select "Confirm bottles removed".
7. Replace the bottles previously removed with fresh reagent, standard solution or cleaning
agent.
Reagents must be prepared according to the instructions for mixing a reagent.
8. Reconnect the hoses to the Liquidmanager.
9. Select "Menu/Operation/Maintenance/Bottle change mode/Bottle insertion/Bottle
selection".
10. Select all the bottles that you replaced and press "OK" to confirm.
11. Select "Confirm bottles inserted".
12. If bottle level monitoring is enabled (Menu/Setup/Analyzer/Extended setup/Diagnostics
settings/Bottles), you can reset it under "Menu/Operation/Maintenance/Bottle change
mode//Bottle insertion/Reset filling levels".
13. The system must be calibrated after a replacement routine. Select "Menu/Operation/Manual
operation/Determine calibration factor".
14. After the calibration return to "MODE/Continue automatic mode" or "MODE/Start automatic
mode", to start normal measurement.
Replacement interval for CA80AM with cooling module:
Reagent CY80AM
Every 3 months
Standard solution CY80AM Every 6 months (typically; depending on the calibration interval)
Cleaner CY800
Every 6 months (typically; depending on the cleaning interval)
Replacement interval for CA80AM without cooling module:
Reagent CY80AM
every 6 weeks (typically, depending on temperature and
measuring range)
Standard solution CY80AM Every 6 months (typically; depending on the calibration interval)
Cleaner CY800
Every 6 months (typically; depending on the cleaning interval)