Liquiline System CA80AM
Maintenance
Hauser
169
12.5 Changing the filter mats
You need the following parts:
• Filter mats (part of the CAV800 maintenance kit)
1. Open and remove the fan screen on the right and left underside of the analyzer.
2. Remove the used filter mats and replace them with new ones from the maintenance kit.
3. Reattach the fan screens.
4. Select "Menu/Operation/Operating time filter mats/Reset"
12.6 Changing the dispensers
You need the following parts:
Interval at which the filter mats need to be cleaned:
• Depends on the application; recommendation: every 6 months
Interval at which the filter mats need to be replaced:
• Depends on the application; recommendation: every 12 months
Interval at which the dispensers need to be replaced:
• 10 ml dispenser s (S7 =
standard)
every 180 days
• 10 ml dispensers (S2 = waste)
• 2.5 ml dispensers (S3 = hose R2;
S4 = hose R1)
• 10 ml dispensers with adapter, mounted
Part of the CAV800 maintenance kit
• 2.5 ml dispensers with adapter, mounted
• 1 pair of gloves that are resistant to the reagents used