15.1.7 Cable specification (for optional sensors with Memosens technology)
Cable type
Memosens data cable CYK10 or fixed sensor cable, each with cable end sleeves or M12
round-pin connector (optional)
Cable length
Max. 100 m (330 ft)
15.2 Output
15.2.1 Output signal
Depending on version:
• 2 x 0/4 to 20 mA, active, potentially isolated
• 4 x 0/4 to 20 mA, active, potentially isolated
• 6 x 0/4 to 20 mA, active, potentially isolated
Modbus RS485
Ethernet and Modbus TCP
15.2.2 Signal on alarm
Adjustable, as per NAMUR Recommendation NE 43
• In measuring range 0 to 20 mA:
Error current from 0 to 23 mA
• In measuring range 4 to 20 mA:
Error current from 2.4 to 23 mA
• Factory setting for error current for both measuring ranges:
21.5 mA
Signal coding
EIA/TIA-485
Data transmission rate
2400, 4800, 9600, 19200, 38400, 57600 and 115200 Baud
Galvanic isolation
Yes
Bus termination
Internal slide switch with LED display
Signal coding
IEEE 802.3 (ethernet)
Data transmission rate
10 / 100 MBd
Galvanic isolation
Yes
IP address
DHCP or configuration via menu