Technical data
Liquiline System CA80AM
192
Hauser
15.5 Protocol-specific data
15.5.1 Modbus RS485
15.5.2 Modbus TCP
15.5.3 EtherNet/IP
6)
Protocol
RTU / ASCII
Function codes
03, 04, 06, 08, 16, 23
Broadcast support for function codes
06, 16, 23
Output data
16 measured values (value, unit, status), 8 digital values (value, status)
Input data
4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features
Address can be configured via switches or the software
TCP port
502
TCP connections
3
Protocol
TCP
Function codes
03, 04, 06, 08, 16, 23
Broadcast support for function codes
06, 16, 23
Output data
16 measured values (value, unit, status), 8 digital values (value, status)
Input data
4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features
Address can be configured via DHCP or the software
6)
available from software version 01.05.04
Protocol
EtherNet/IP
ODVA certification
yes
Device profile
Generic device (product type: 0x2B)
Manufacturer ID
0x049E
h
Device type identification
0x109C
h
Polarity
Auto-MIDI-X
Содержание Liquiline System CA80AM
Страница 2: ......