714 Filter mats change
M
On
Off
Filter mat change necessary
The limit value for operating hours has been
exceeded
-> Replace the filter mats and reset the operating
hours counter in the Diagnostics menu
715 Calibration expired
M
On
Off
The validity of the last calibration has expired.
The date of the last calibration is too far in the
past.
Measurement can still take place.
Possible reasons:
• Manual intervention has prevented automatic
calibration
-> Calibrate analyzer manually
-> Check device configuration
716 Calibration expired
S
On
Off
Calibration has failed or is not reliable
Possible reasons:
• Stability criteria are not met
-> Check the configuration and perform a manual
calibration again
-> Contact the Service Department
717 Photometer LED
F
On
On
Photometer defective
Possible reasons:
• No LED voltage
• No LED current
-> Contact the Service Department
718 Cuvette covering
M
On
Off
Photometer test
High level of contamination - reliable
measurement will not be possible soon
1. Clean analyzer with manual cleaning process
2. Contact the Service Department
3. Replace photometer
726 Liquids warning
M
On
Off
Consumable liquids, warning
Level monitoring or shelf life
Measurement can still take place.
Possible reasons:
• The level of one or more liquids is low or the
expiry date of one or more liquids is
approaching.
Top up/replace liquids in question and reset
counter in Diagnostics/Operating time
information
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current