Liquiline System CA80AM
Operation
Hauser
81
Data management
Firmware update
Please contact your local sales office for information on firmware updates available for your
controller and its compatibility with earlier versions.
The
current firmware version
of the analyzer, FXAB1 control module, photometer and sample
preparation 1 can be found at: Menu/Diagnostics/System information/Software version.
First save your current setup on an SD card since a firmware update overwrites your
settings with the factory settings. After updating the firmware, you can restore your setup
by uploading it from the SD card.
To install a firmware update, you must have the update available on an SD card.
1. Insert the SD card into the controller card reader.
2. Go to: Menu/Setup/General settings/Extended setup/Data management/Firmware update.
The update files on the SD card are displayed.
3. Select the desired update and select yes when the following question is displayed: The
current firmware will be overwritten. After this the device will reboot. Do you want to
proceed?
The firmware is loaded and the device is then started with the new firmware.
Updating the measurement parameter lists
The measurement parameter lists contain analyzer-specific parameters and constitute a
part of the software.
Please contact your local sales office for information on updates that are available for the
measurement parameter lists for your analyzer and their compatibility with earlier versions and
with firmware versions.
The
current version
of the analyzer's measurement parameter list can be found at:
Menu/Diagnostics/System information/MPL version.
To install a MPL update, you must have the update available on an SD card.
1. Insert the SD card into the analyzer card reader.
2. Go to: Menu/Setup/General settings/Extended setup/Data management/Meas. param. list
update.
The update files on the SD card are displayed.
3. Select the desired update and select yes when the following question is displayed: The
current parameter configuration will be overwritten. After this the device will reboot. Do
you want to proceed?
The measurement parameter list is loaded and the device is then started.