EM TEST (Switzerland) GmbH
PFM 200N100.1
User’s Guide
Ver. 1.12
14 / 43
4.2.
Back Panel Elements
Figure 4.2
1
PF1 IN, DUT power input High
2
PF1 IN, DUT power input Low
3
Reference Earth
4
Cooling Fan
5
Frame Bus Interface
6
Mains Power Input and Fuse
7
Power Switch ON/OFF
1
PF1 IN, DUT Power Input Low (VDS)
Input connector (4 mm and 6 mm) for DC power for the DUT (Low).
The generator must be connected to the reference earth of the test setup.
2
PF1 IN, DUT Power Input High (VDS)
Input connector (4 mm and 6 mm) for DC power for the DUT from an external source (High)
Max. DC Voltage ±100V
- 6 mm connector 100 A
- 4 mm connector 32 A max.
3
Reference Earth
Earth connector with an M5 screw terminal and 4mm banana connector. The generator must be connected
to the reference earth in the test setup.
4
Cooling Fan
Cooling unit. At least 20cm around should be allowed for proper ventilation.
5
Frame Bus Interface
Interface for controlling the PFM 200N100.1 from an AutoWave. The output of the Frame Bus must be
terminated with the Framebus Terminator part no. 101732.
6
Mains Input and Fuse
The mains input has an integrated filter and fuse for the generator using (2x 2 AT) fuses. The generator power
supply is a universal power supply with a range of 100 V - 265 VAC, 50/60 Hz
7
Power Switch
Power ON/OFF Switch