7.3.1. Reference Measurement with the AutoWave.control Software
The reference measurement can be selected by
selecting Edge Verification in the software.
There is a test window with where you can call up
the settings for the reference measurement.
1
Edge Verification
Choose this option to perform the reference
measurement.
2
Test Window for the reference measurement
Figure 7.10: Test Window for Reference Measurement of the von
Power Lines
The load resistor must have the following characteristics to achieve the required pulse parameters:
-
Low Inductive Component
-
Shortest possible cables (a few cm) to the PFM output
-
Probe must be as near as possible to the load resistor or, as with the CA LV 124-P, using a BNC
connection directly on the load.
Window to Control the Reference
Measurement
Select verification box
Select the calibration load
1
CA LV124-P1R
1
Ω
load resistor for max. 12V
2
CA LV124-P100R
100
Ω
load resistor for max. 100V
3
Other
Other load resistors
4
Set the Verification Voltage
The software sets the voltage when using the
EM Test loads automatically. Carefully
observe the power of the load resistors.
5
Start the Test
6
The
OK
closes the window
Figure 7.11: Reference Measurement Test Window
Continuous Verification
1
Continue
Continue the verification after a break
2
Break
Pause
3
Stop
Stops the verification