EM TEST (Switzerland) GmbH
PFM 200N100.1
User’s Guide
Ver. 1.12
13 / 43
4.
Functions and Operation of the PFM 200N100.1
4.1.
Front Panel Elements
Figure 4.1
1
Test enable button
"TEST ON”
2
LED Power ON
3
LED Fail
4
Data line inputs L1
– L16
5
Data line outputs L1
– L16
6
Monitor BNC port
7
Trigger OUT BNC port
8
Trigger IN BNC port
9
DUT Test Supply LOW
10
DUT Test Supply HIGH
1
Test Enable Button
“TEST ON”
Pressing the "TEST ON" button enables testing. The LED serves also as an optical indication when the switch
switches: the DUT power will be provided when the button is lit.
2
LED Power ON
This LED will light when the switch on the back of the device is switched on, indicating that the PFM is ready.
3
LED Fail
This LED will light when the PFM 200N100.1 detects an error. An error message will be displayed in a window in
the AutoWave.control software.
4
Data Line Inputs L1
– L16
16 pin PHOENIX, DFMC connector with 1.5 mm2, push-in locking connections.
Upper row: switched channel L1 to L16 (input)
Bottom row: return lines for the upper L1 to L16 switches
5
Data Line Outputs L1
– L16
16 pin PHOENIX, DFMC connector with 1.5 mm2, push-in locking connections.
Upper row: switched channel L1 to L16 (output)
Bottom row: return lines for the upper L1 to L16 switches
6
Monitor
Monitoring output for the DUT voltage at the test supply output using a differential measurement
Ratio 1:20, Amplitude 10 Vpp, ± 5 V
7
Trigger OUT, Oscilloscope Trigger
BNC trigger output (12 V pos. edge) for external measurement devices
8
Trigger IN
Trigger input to control the trigger the switch.
Signal (5 V neg. edge), delay ca. 100 µs
9
DUT Test Supply LOW
DUT output LOW with 4 mm (male) / 6 mm (female) connectors
10 DUT Test Supply HIGH
Switch DUT voltage output LOW 4 mm (male) / 6 mm (female) connectors